Pixel Array Size Needed to Replace Photomicrographs on Film
Introduction
Some of you may have read previous articles by me in “Microscopy Today” and know that I am a strong proponent of digital imaging for photomicrographs and photomacrographs. I require that the digital images match the resolution and field size of traditional film images, with the 4X5 Polaroid film most commonly used in metallurgical laboratories where I worked before early retirement. My development of an affordable universal student microscope for home microscopy resulted in a need for color digital images that would meet my requirements at a reasonable cost. This need was met by recording on 35 mm film with subsequent scanning of selected film images to a Kodak Master Photo CD. More recently I purchased a CanoScan FS2710 film scanner for digitizing my family’s collection of slides and negatives for saving on CDs. This article will review how I arrived at my requirements and how well they were met by digital cameras and by film scanning.
Microscope Resolution
The definition of the spatial resolution of the microscope begins with the diffraction pattern image of two nearly adjacent point sources of light formed through a circular aperture. The point sources are separated in object space so that the first minimum of one pattern coincides with the central maximum of the other diffraction pattern. This is the Rayleigh criterion of resolution. The image of these two point sources is the sum of the two pattern intensities as shown in Figure 1. Recording a straight line series of overlapping Airy discs at the Rayleigh limit with a CCD would require at least two pixels per Airy disk with the pixels centered on the central maximum and first minimum corresponding to a pixel size of one fourth the Airy disc diameter. The delta Y separation in the image can be related back to the separation of the point sources in object space through the application of Abbe’s sine condition for a lens free of spherical aberration; this is the objective lens resolution given in Figure 2.
|
Field Size
The recording of photomicrographs requires consideration of the field size limitations of the microscope. The compound microscope optics produce an intermediate image which was conventionally enlarged for viewing with a 10X eyepiece and projected with 10X enlargement for recording on 4X5 film or with 2.5X enlargement for recording on 35mm film. Until recently the intermediate field size was limited to about 18mm. Modern research microscopes may now have an intermediate field size as large as 26mm in diameter as shown in Figure 3. The size of the rectangle in the intermediate image enlarged to form the photomicrograph is important in determining the physical size of the CCD array and relay lens magnification for digital recording. The intermediate field size traditionally recorded on 4X5 film is an 8.9X11.4mm rectangle as shown in Figure 3. Recording a 17.8X17.8mm square format image from the 26mm intermediate image diameter would make good use of the modern optics when necessary.
Final Print Resolution
Resolution of the final photomicrograph is rarely mentioned in quantitative terms. The viewing and recording optics of the light microscope have been consistent with Abbe’s definition of useful magnification, which is related to the resolving ability of normal vision when the photomicrograph is viewed from 250mm. Images from the lower power objectives generally meet an Abbe magnification criterion of 500X the NA of the objective. An Abbe criterion of 1000X the NA is commonly used with the highest power objectives so that the finest detail resolvable by the objective can be easily seen when higher NA is not available. Images from the lower power objectives place more demand on the resolution of the film or CCD, as indicated in the analysis given in Figure 4. Assuming a traditional 10X enlargement of the intermediate image to achieve photomicrograph resolution of between 6 and 3 line pairs per mm, the intermediate image resolution ranges between a corresponding 60 and 30 line pairs per mm.
|
Pixel Array Size Requirement
A minimum pixel array size that is needed to achieve 6 line pairs per mm resolution in the image area of a 4X5 Polaroid film print is derived in Figure 5. The calculation agrees well with the now widespread use of 1280X1024 through 1600X1200 pixel CCD cameras for scientific imaging. The calculation assumes that the three-color values for each pixel are not interpolated as with consumer product cameras having a color mosaic filter pattern over the pixels, which reduces spatial resolution by at least 30%. Non-interpolated pixel values can be obtained by many approaches. The most common non-interpolated color capture is by sequential capture of red, green and blue images using a color filter wheel or tunable liquid crystal filter. Film scanners and some color digital cameras record the image by scanning the optical image with a red, green and blue filtered trilinear array.
Resolution Test Patterns
Measuring the resolution capability of a digital camera or film scanner for photomicrographs ideally requires a microscope resolution test slide. The lowest magnification objective is commonly a 5X 0.10 NA (4X 0.10 NA used with a 1.25X tube factor in some Zeiss microscopes). This lens should resolve 300 line pairs per mm in object space forming an intermediate image with 60 line pairs per mm resolution. A 3X3” chrome on glass resolution test chart with patterns from ¼ line pair per mm to 600 line pairs per mm is available and quite costly. I previously used this test chart with a 0.10 NA Zeiss objective and verified the theoretical resolution with both the MegaPlus 1.6i/AB digital camera and recording on Polaroid Type 55 film. Low cost test patterns with a small field size with line patterns covering a range of 300 to 3000 line pairs per mm could be made by electron lithography but are not available because of lack of market demand. An alternative test method is to use a high resolution macro lens on the camera to image a pattern like the NBS Microcopy Resolution Test Chart in the 89X114 mm field size of a 4X5 Polaroid film print for a 1280X1024 pixel CCD sensor (89X133 mm field for a 1534X1024 CCD sensor or film scan of 35 mm format) as done for Figures 6 & 7.
FIGURE 6 Comparison of digital camera resolution with Kodak Photo CD resolution using the highest resolution portions cropped from the original images. |
FIGURE 7 Resolution test results for the Canon CanoScan FS2710 film scanner. |
Digital Camera and Kodak Photo CD Resolution
The test result for the Mavica camera demonstrates the significant loss in spatial resolution due to the color mosaic filter over the CCD sensor when compared with the grayscale capture by the MegaPlus camera. The pattern results with the MegaPlus camera demonstrate that line patterns oriented orthogonal to the pixel array are more difficult to record without an alias line problem. The alias lines are a moiré fringe effect from close alignment of the pixel spacing with the line patterns in the optical image when the spatial frequency of the line patterns approach the pixel spacing. The moiré effect was accentuated with the MegaPlus results because the original images were high pass filtered to improve the apparent resolution. The position of the test pattern during imaging was purposely adjusted to avoid this effect for the orthogonal orientation of the 5.6 line pairs per mm pattern. The orthogonal patterns from the Photo CD scans show less evidence of this effect, perhaps mainly because they were not sharpened. The solution to the alias line problem is to capture the same image with a larger number of pixels. Since my home microscope and the older Zeiss Universal microscope I used before retirement have a maximum intermediate image field diameter of 18 mm, a 1534X1024 pixel array meets my requirements for resolution and field size. The 3000X2000 pixel resolution result indicates that the 35 mm film to digital method using this Kodak Photo CD file size cropped to 2048X2048 pixels should be adequate to cover an 18X18 mm portion of a 26 mm intermediate image field diameter for those fortunate enough to own a modern research microscope. The 11 line pair per mm pattern just resolved in the 3000X2000 pixel file corresponds to 40 line pairs per mm in the film image. The blurring effect of limited film resolution is showing up at this higher spatial frequency.
Canon Film Scanner Resolution
Film scanners for use with fast PC’s with 8 megabyte graphics cards, 1600X1200 pixel monitors and CD writers are now affordable for home use. I have a large collection of family 35-mm slides and negatives and wanted to digitize them with storage on CD’s for distribution to the younger members of the family so their past will not be lost. The cost savings of owning my own film scanner versus having the film images scanned to Kodak Master Photo CD’s justified purchase of a Canon CanoScan FS2710 film scanner. A maximum resolution scan for this scanner is 3888X2720 pixels. This pixel array size is not adequate to record the finest spatial frequencies in very high quality slides and negatives, but should achieve about 52 line pairs per mm resolution. Some 35 mm film images exceed 63 line pairs per mm resolution. The need to digitize the highest spatial frequencies on 35 mm film was met by the Kodak Professional Photo CD 6000X4000 pixel scans. Film scanners with this resolution are now available, but quite expensive for home use. Owning the CanoScan film scanner is a big advantage for my low budget, home microscopy. I can record on 35 mm color negative film and use locally available 1 hour processing of the film prior to scanning it at home. Scanning the negatives rather than photographic prints from the negatives avoids the problem of variable print quality, which the mass market probably would not notice. The disadvantage is that I have to bracket my exposures to be sure of a good film image to scan. Naturally I have compared the resolution of the Canon scanner, using the same resolution test slide, with the Kodak Photo CD results in Figure 6. The finest pattern on the NBS test chart just resolved on the film is 14 line pairs per mm pattern corresponding to 51 line pairs per mm on the slide. PhotoShop LE comes with the scanner software and was used to downsize the large file size to 1940X1360 and 1534X1024 pixel files. A photomontage of the finest patterns in these files is shown in Figure 7. It is important to note that the other than 2:1 downsizing used to obtain the 1534X1024 pixel file introduces noticeable alias line artifacts less evident than in the Photo CD result shown in Figure 6. The 30% loss in spatial resolution resulting from use of a CCD camera with a Bayer color mosaic filter, such as the Nikon Coolpix 995, can be compensated by doubling the number of pixels used to capture the same image field; this is shown in Figure 8. The bicubic resampling procedure of PhotoShop can then be used to maintain the spatial resolution using half the number of pixels. This step is equivalent to reducing the empty magnification.
FIGURE 8 Resolution Test of the Nikon Coolpix 995 Camera. |
Summary
Film-to-digital and direct digital capture, with a digital camera, can clearly be acceptable methods for obtaining high quality digital photomicrographs that match the field size and resolution of traditional photomicrograph prints. Unfortunately there are no national or international standards dealing with this important transition in technology. The Photographic and Imaging Manufacturers Association (PIMA) is participating in an ISO TC42 effort to develop such standards. Their preliminary efforts concentrate on using a resolution test pattern of their own design and alias effects are considered. Alias effects are inherent in the digital capture, but should not be worsened in subsequent image downsizing and desktop publishing. Understanding the operating principles of the light microscope and using them to obtain an optimum optical image for recording is vital for quality digital images. Unfortunately the users who think software can “fix” the resulting defects in their images do not appreciate this requirement.
Reference
Clarke, T. M., “Digital Imaging in the Materials Engineering Laboratory,” THE MICROSCOPE, 1998, 46(2), 85-100.
Clarke, T. M., “Building an Affordable Universal Student Microscope,” THE MICROSCOPE, 2000, 48(1), 19-39.
K. D. Moller, OPTICS, University Science Books: Mill Valley, Ca., 1988. Edwards, Parulski, & Holm, “Setting Standards–Developing Standards in Electronic Imaging”, PEI, February 1998, 48-52.
Comments
add comment